Classification: | Imaging Diagnostic Equipment |
---|---|
Type: | X Ray Equipment |
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Part Name
|
Model
|
Qty
|
Note
|
Control Console
|
KZT10
|
1 unit
|
YHX
|
HF HV generator
|
FSQ80
|
1 set
|
YHX
|
X-ray tube assembly
|
E7254FX
|
1 unit
|
Canon (Toshiba)
|
Collimator
|
XSQ30
|
1 unit
|
YHX
|
Suspension frame
|
ZDT40
|
1unit
|
YHX
|
Photography Table
|
SYC30
|
1 unit
|
YHX
|
Bucky Stand
|
XPJ40
|
1 unit
|
YHX
|
Flat Panel Detector
|
YHX-1717V
|
1 set
|
YHX
|
Image acquisition Workstation
|
DRA20
|
1 set
|
YHX
|
Monitor
|
P1917S
|
1 set
|
Dell
|
Item
|
Content
|
|
Technical Parameter
|
||
Power Supply
|
Voltage
|
|
380V±38V
|
||
|
Frequency
|
|
50Hz±1HZ
|
||
|
Capacity
|
|
≥105kVA
|
||
|
internal resistor
|
|
≤0.11Ω
|
||
Generator
|
Power Output
|
|
80KW
|
||
|
Inverter Frequency
|
|
440 KHz
|
||
|
Tube Voltage
|
|
40kV-150kV
|
||
|
Tube Current
|
|
10mA-1000mA
|
||
|
Exposure time
|
|
1.0ms-10000ms
|
||
|
mAs
|
|
0.1-1000mAs
|
||
Digital Controlled X-ray Tube
|
Tube Focus:Large Focus/Small Focus
|
|
1.2mm /0.6mm
|
||
|
Input Power
|
|
Big focus 102kW / Small Focus 40kW
|
||
|
Anode Thermal capacity
|
|
285KJ (400kHu)
|
||
|
Anode heat dissipation rate
|
|
1180W (1664kHU/min)
|
||
|
Tube thermal capacity
|
|
950KJ (1339kHu)
|
||
|
Rotary anode speed
|
|
9700rpm
|
||
Suspension frame
|
lateral movement
|
|
≥2300mm
|
||
|
longitudinal motion
|
|
≥2100mm
|
||
|
up-and-down motion
|
|
≥1450mm
|
||
|
X-ray rotate around horizontal axis angle
|
|
±90°
|
||
|
X-ray rotate around vertical axis angle
|
|
360°
|
||
Radiography table
|
Longitudinal movement of table surface
|
|
900mm
|
||
|
Lateral movement of table surface
|
|
220mm
|
||
|
Table surface height from the ground
|
|
500mm-720mm ( lifting range:220mm)
|
||
|
Capacity
|
|
135KG
|
||
Flat Panel Detector
|
Vision size 427(H)×427V)
Pixel Matrix 3072(H)×3072V) Pixel gap139μm Limiting resolution3.1 lp/mm typ A / D transform14bit Energy range40 - 150 kVp Power input DC 24V |
|
|
||
Work Station
|
CPU System
|
Brand: DELL OPTIPLEX 5050MT
ProcessorCore™ i5-6500(3.20GHz) RAM4GB DDR3L Hardware1T SATA (7200 rpm) CD-driversuit MT's 16X variable-speed DVD + / - RW With functions of double written Network Card: Broadcom NetXtreme 10/100/1000 PCIe High display card: AMD RADEON R5 430,2GB, PCIE seriel port and parallel port, all height |
|
||
|
Color LCD Monitor
|
19 inches Dell
|
|
||
|
Work Station Software
|
Basic operation: change control console password, edit ID, acquisite images.
|
|
||
|
|
Additional operation:Add new check, edit present checking info, add new position, change image aquisition order, multi-checking
agreement chance, manual adjustment of exposure parameters, automatic exposure control mode, focus choice, patient body-type choice, tube capacity check, ESA curve choice, image cutting, note added (sent to DICOM workstation) , mark on images, rotate or overturn, full-size image observation, check patient info and dose info, accept or refuse images. |
|
||
|
|
Image management: change order, patient basic info editing, inquiry history images, resend history images, re-print history
images, check images mark info, review history images, manage refused images, space reclaimed, image protection, and manual image deletion etc. |
|
||
|
|
System managementID edition, change ID password, ED refrigeration set, statistics info checking, detector calibration, equipment
control, output order management, image measurement. |
|
||
|
|
|
|