Application Scope
The product is applicable to disinfect and sterilize medical devices, dressings, glass wares and hydroponics medium etc in the steam method.
Features and parts function introduction
1. The sterilizer body is a closed double-layer structure consisting of a sterilization chamber and a steam jacket. When the steam enters and fills the jacket, the walls of the sterilization chamber are also preheated. A solenoid valve is arranged between the sterilization chamber and the jacket, and the pressure steam enters the sterilization chamber through the main valve through the jacket, so that the articles are wetted and heated, and the purpose of sterilization and sterilization is achieved under the action of humidity and heat. At the same time, it can meet the requirements of preheating, drying and continuous operation of the articles to be sterlized
2. The jacket of the sterilization chamber is covered with a casing and filled with insulation wool with good thermal insulation properties, which is conducive to saving energy and avoiding excessive rise in ambient termperature
3. The opening and closing of the sterilization chamber door adopts the radiation rod rotary type, and both are equipped with two-stage self-locking and interlocking safety devices. When closing the door, firstly pull the safety tight hand wheel, the adiation rod to the fully engaged position, and then turn The hand wheel, the center turntable makes an axial entry, and the center turntable is self-locked by the positioning pin. The door of the sterilizer is equipped with a safety interlock. Under normal working conditions, steam should not enter the sterilization chamber when the door is not locked. It is ensured that the pressure of the sterilizer door in the sterilization chamber has been safely released, and the safety interlock device can be opened after being disengaged, otherwise the door should not be opened. The interlock device exits but the turntable is still locked by the positioning pin, the hand wheel can be reversed, the center turntable can only be axially withdrawn, the door seal gradually fails, and the remaining steam is released to zero, that is, the pressure is first released, and then the door is opened. When the hand wheel is retracted to the outermost position, the safety tightening hand wheel can be pulled out to exit the radiation rod to open the door.
4. The sterilizer is provided with a program function, and the air and condensed water in the sterilizer are automatically discharged during the sterilization process, so that the sterilization temperature in the sterilization chamber tends to be balanced
5. A device equipped with a steam pressure to maintain the voltage regulation, the pressure control is set in the range of 0.05 to 0.22 MPa, and the corresponding temperature is 110 to 134 °C. The operating temperature pressure is selected via the adjustment button on the control panel.
6. With a timer device, the time is set from 0 to 99 minutes, and the working time is selected by the adjustment button on the control panel
7. It is equipped with a temperature control failure protection device, and the top of the sterilizer is equipped with a micro-opening spring safety valve. When the steam pressure in the jacket exceeds the maximum allowable range, it can automatically open and release the super Pressurized steam; when the steam pressure in the container returns to normal, it can be automatically closed. Make sure the safe use, the working pressure range of the safety valve is 0.22~0.24MPa.
8. With manual maintenance drain valve, when the evaporator of this equipment needs to be cleaned and drained, the manual valve at the bottom left of the equipment can be opened to drain the water tank
9. equipped with solenoid valve:
* Inlet valve, 2 only press the start button program to automatically enter the water to the high water level and then close, the vacuum pump enters the water,
* Supply valve, the sterilization will open automatically,
* Exhaust valve, power on and press the start button program to automatically vent, automatically open and close during the sterilization process.
* Sterilization valve, which automatically opens during sterilization.
* The vacuum pump is automatically opened when the vacuum pump