Instrument classification
Properties
Medical Materials & Accessories
TECHNICAL SPECIFICATIONS
Operating Principle: Partial pressure electrochemical
Output: 9 mV to 13 mV in 210 mBar O2
Measurement Range: 0 mBar to 1500 mBar O2
Baseline Offset: <200 μV
Temperature Compensation: <2% O2 equivalent (0°C to 40°C)
Connector: 0.141 in/3,5 mm mini phone jack
Housing Material: White ABS
Weight: 42 g (nominal)
Orientation: Any
Typical Applications: Critical care anesthesia
Operating Temperature Range: -20°C to +50°C
Operating Pressure Range: 0.5 Bar to 2.0 Bar
Operating Humidity Range: 0% to 99% RH non-condensing